device · product 1 of 1
AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI: 15099590364083 , Model #ODL25255
Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .
These labels are deterministic app interpretations, not FDA categories.
Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .
Code information
All units containing software Version(s) 4.1.1 and up
Distribution pattern
US: NY, VA OUS: Austria Belgium Czech Republic France Germany Israel Italy Lithuania Mayotte Romania Slovakia South Africa Spain Sweden Switzerland United Kingdom