device · product 1 of 1
Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298
- Recall number
- Z-1105-2020
- Initiated
- December 04, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 286 Monopolar curved scissors
App-derived interpretation
This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.
Code information
Serial Number/Manufacturing Lot: N11181003, N11181004
Distribution pattern
US National Distribution Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming OUS:None