Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 84054

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 08, 2019
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Teleflex Medical

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

Z-0688-2020
Recall number
Z-0688-2020
Initiated
October 08, 2019
Classification
Class II
Status
Ongoing
Recalling firm
Teleflex Medical
Quantity
526,000 total

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
packaging may compromise the sterility

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.

Code information

All

Distribution pattern

US nationwide distribution.

device · product 2 of 3

Arrow EZ-IO Intraosseous Vascular Access System 25mm Needle+Stabilizer Kit, Product Code 9001P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

Z-0689-2020
Recall number
Z-0689-2020
Initiated
October 08, 2019
Classification
Class II
Status
Ongoing
Recalling firm
Teleflex Medical
Quantity
526,000 total

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
packaging may compromise the sterility

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.

Code information

All

Distribution pattern

US nationwide distribution.

device · product 3 of 3

Arrow EZ-IO Intraosseous Vascular Access System 45mm Needle+Stabilizer Kit, Product Code 9079P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

Z-0690-2020
Recall number
Z-0690-2020
Initiated
October 08, 2019
Classification
Class II
Status
Ongoing
Recalling firm
Teleflex Medical
Quantity
526,000 total

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
packaging may compromise the sterility

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.

Code information

All

Distribution pattern

US nationwide distribution.