Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 84112

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 25, 2019
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Nidek, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

NIDEK Specular Microscope, Model CEM-530 - Product Usage: The device provides noncontact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images.

Z-0771-2020
Recall number
Z-0771-2020
Initiated
February 25, 2019
Classification
Class III
Status
Terminated
Recalling firm
Nidek, Inc.
Quantity
13 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)

Code information

Serial numbers 120080, 120081, 120086, 120087, 120088, 120090, 120091, 120092, 120093, 120094, 120095, 120096, and 170001.

Distribution pattern

US Nationwide distribution in the states of AZ, CA, IL, KS, MD, NJ, NY, PA, SC, and VA, as well as PR.