device · product 1 of 1
XEN Gel Stent, REF 5513-001(US Model), Sterile, RX Only, (01)10888628032439 ; Other Model Numbers: 5507-001(Global Model) and 5517-001(Australia model) Implant for glaucoma treatment.
- Recall number
- Z-0598-2020
- Initiated
- October 30, 2019
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Allergan PLC
- Quantity
- 145,711 Glaucoma Treatment Systems
App-derived interpretation
foreign body
Official device-enrichment evidence · Sourced
Vendor change control
Inspect official wording and provenance
Reason for recall
Residual polishing compounds, that are used in the needle sleeve manufacturing process, may be present on the injector needle. The residual polishing compounds could transfer to the patient's eye during procedure possibly resulting in irritation, inflammation, local allergic reaction/ hypersensitivity, iritis, uveitis/sterile endophthalmitis or an intraocular foreign body.
Code information
All Lots within Expiry: US Lot codes: 61650, 61657, 61779, 61780, 61825, 61883, 61884, 61906, 61911, 62578, 62608, 62703, 62812, 62874 and 62941 OUS Lot codes: 61566, 61580, 61626, 61642, 61685, 61846, 61847, 61955, 61966, 62008, 62301, 62053, 62066, 62108, 62130, 62263, 62297, 62318, 62380, 62538, 62636, 62678, 62683, 62719, 62745, 62749, 62769 and 62890
Distribution pattern
Worldwide distribution. US nationwide, countries of AE, AT, AU, BE, BG, CA, CH, CN, CZ, DE, DK, ES, FI, FR, GB, GR, HK, HR, IE, IL, IS, IT, KR, LT, MT, MU, MX, MY, NA, NL, NO, NZ, PH, PL, PT, RO, RS, SA, SE, SG, SI,SK, SM,TH,TR, TW, VE, and ZA.