Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 84227

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 05, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Avanos Medical, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology, SAP Finished Product Code 109382803; 3. REF 202, Elbow, SAP Finished Product Code 109382903; 4. REF 195-5 Y-Adapter, SAP Finished Product Code 109838302.

Z-0788-2020
Recall number
Z-0788-2020
Initiated
November 05, 2019
Classification
Class II
Status
Terminated
Recalling firm
Avanos Medical, Inc.
Quantity
114,300 units (37,880 US, 76,420 OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Manufacturing material removal

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation.

Code information

REF 195 Lot Code / Expiration Date: M18115T401/2023-APR, M18121T402/2023-MAY, M18128T404/2023-MAY, M18143T401/2023-MAY, M18149T403/2023-MAY, M18156T402/2023-JUN, M18170T402/2023-JUN, M18177T402/2023-JUN, M18192T402/2023-JUL, M18206T401/2023-JUL, M18220T402/2023-AUG, M18247T402/2023-SEP, M18254T405/2023-SEP, M18282T404/2023-OCT, M18311T401/2023-NOV, M18318T402/2023-NOV, M18345T403/2023-DEC, M1834T403/2023-DEC, M18358T403/2023-DEC, M19008T401/2024-JAN, M19058T402/2024-FEB, M19078T403/2024-MAR, M19086T401/2024-MAR, M19098T403/2024-APR, M19134T404/2024-MAY, REF 201 Lot Code / Expiration Date: M18044T405/2023-FEB, M18311T402/2023-NOV, M18358T404/2023-DEC, M19037T402/2024-FEB, M19065T402/2024-MAR, M7172T402/2022-JUN. REF 202 Lot Code / Expiration Date: M18143T403/2023-MAY, M18192T401/2023-JUL, M18254T404/2023-SEP, M18274T401/2023-OCT, M18296T405/2023-OCT, M18318T401/2023-NOV, M19072T401/2024-MAR, M19142T401/2024-MAY, M7346T403/2022-DEC. REF 195-5 Lot Code / Expiration Date: 20000095/2024 - JUL, M18274T402/2023-OCT, M18290T402/2023-OCT, M18296T406/2023-OCT, M18311T403/2023-NOV, M18324T401/2023-NOV, M18324T402/2023-NOV, M18332T401/2023-NOV, M18339T401/2023-DEC, M18345T404/2023-DEC, M18352T402/2023-DEC, M19024T402/2024-JAN, M19030T401/2024-JAN, M19051T401/2024-FEB, M19059T401/2024-FEB, M19065T401/2024-MAR, M19072T402/2024-MAR, M19079T401/2024-MAR, M19085T402/2024-MAR, M19098T404/2024-APR, M19099T401/2024-APR, M19105T403/2024-APR, M19120T403/2024-APR, M19128T401/2024-MAY, M19128T403/2024-MAY, M19135T401/2024-MAY, M19142T402/2024-MAY, M19149T405/2024-MAY, M19162T403/2024-JUN, M19177T402/2024-MAY, M19184T401/2024-JUL.

Distribution pattern

Nationwide. Internationally distributed to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Croatia, Czech Republic, Denmark, France, Germany, Greece, Honduras, India, Ireland, Israel, Italy, Korea, Republic of, Lebanon, Mexico, Nepal, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.