openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804
It was reported that the user's low alarm feature on the iOS application were not properly alerting users when the user has enabled the Alert Schedule feature more than 30 days after installing the iOS application.
These labels are deterministic app interpretations, not FDA categories.
It was reported that the user's low alarm feature on the iOS application were not properly alerting users when the user has enabled the Alert Schedule feature more than 30 days after installing the iOS application.
Code information
Software Number: SW11677 Software Version: 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804
Distribution pattern
US Nation-wide OUS: Austria; Belgium; Canada; Czechia; Denmark; Finland; Germany; Great Britain; Hong Kong; Ireland; Italy; Luxembourg; Netherlands; Norway; South Africa; Spain; Sweden, and Switzerland.