Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.
- Recall number
- Z-1717-2020
- Initiated
- January 21, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Sysmex America, Inc.
- Quantity
- 40 units
App-derived interpretation
When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
Code information
Serial Numbers: 11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14969, 14970, 14972, 14973, 14974, 14975, 14976, 14977, 14978, 14979, 15017, 15018, 15619, 15620
Distribution pattern
USA: FL, VT, IL.