Exact FDA event identifier

Event 84883

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 22, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Braemar Manufacturing, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only

Z-1459-2020
Recall number
Z-1459-2020
Initiated
January 22, 2020
Classification
Class II
Status
Terminated
Quantity
10582

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software. Error

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

Code information

All devices manufactured or distributed prior to 01/01/2020 are affected.

Distribution pattern

Nationwide. No governmental consignees or Canadian consignees. International distribution to Germany, Spain, Switzerland, France, Finland, Vietnam, Israel, France, Hong Kong, Czech Republic, Finland, Belgium, Sweden, Mexico, Lebanon, United Arab Emirates, Bolivia, Italy, and Belgium.

device · product 2 of 2

Braemar Manufacturing LLC DL950 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only

Z-1460-2020
Recall number
Z-1460-2020
Initiated
January 22, 2020
Classification
Class II
Status
Terminated
Quantity
10582

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software. Error

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

Code information

All devices manufactured or distributed prior to 01/01/2020 are affected.

Distribution pattern

Nationwide. No governmental consignees or Canadian consignees. International distribution to Germany, Spain, Switzerland, France, Finland, Vietnam, Israel, France, Hong Kong, Czech Republic, Finland, Belgium, Sweden, Mexico, Lebanon, United Arab Emirates, Bolivia, Italy, and Belgium.