device · product 1 of 4
Transducer S7-3t UDI (01)00884838061439 REF 989605406772 Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70 Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
These labels are deterministic app interpretations, not FDA categories.
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
Code information
All lots.
Distribution pattern
Worldwide distribution - US Nationwide distribution.
device · product 2 of 4
Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound Systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
These labels are deterministic app interpretations, not FDA categories.
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
Code information
All lots.
Distribution pattern
Worldwide distribution - US Nationwide distribution.
device · product 3 of 4
Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
These labels are deterministic app interpretations, not FDA categories.
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
Code information
All lots.
Distribution pattern
Worldwide distribution - US Nationwide distribution.
device · product 4 of 4
Transducer S8-3t UDI (01)00884838067523 REF 989605420183 Can be used with: iE33 Utrasound System - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
These labels are deterministic app interpretations, not FDA categories.
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
Code information
All lots.
Distribution pattern
Worldwide distribution - US Nationwide distribution.