Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 84918

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 17, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Beckman Coulter Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.

Z-1741-2020
Recall number
Z-1741-2020
Initiated
May 17, 2019
Classification
Class II
Status
Terminated
Recalling firm
Beckman Coulter Inc.
Quantity
291 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software design problem

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.

Code information

All serial numbers installed before 5/20/2019: Serials numbers in the U.S.: PW211409002, PW211501005, PW211501006, PW211501007, PW211501008, PW211501013, PW211505001, PW211508003, PW211509002, PW211510001, PW211510002, PW211511001, PW211511002, PW211511003, PW211511005, PW211512001, PW211601001, PW211601002, PW211602002, PW211602003, PW211602004, PW211603001, PW211603002, PW211603003, PW211604001, PW211604002, PW211604005, PW211604006, PW211604009, PW211604010, PW211605001, PW211605002, PW211605004, PW211605005, PW211605006, PW211605007, PW211606005, PW211606007, PW211607001, PW211607004, PW211607005, PW211607009, PW211607017, PW211607018, PW211608001, PW211608006, PW211610002, PW211610006, PW211610007, PW211610008, PW211610011, PW211611001, PW211611002, PW211611003, PW211611004, PW211611005, PW211612003, PW211612006, PW211612007, PW211612008, PW211612010, PW211612013, PW211701001, PW211701010, PW211701013, PW211703004, PW211706008, PW211706009, PW211707005, PW211707006, PW211707007, PW211707009, PW211707011, PW211707012, PW211708008, PW211708009, PW211708010, PW211709004, PW211709011, PW211710001, PW211710009, PW211710010, PW211710012, PW211711002, PW211711004, PW211711007, PW211712001, PW211802013, PW211804005, PW211804006, PW211807006, and PW211807008. Serial numbers OUS: PW201607008, PW211409001, PW211501001, PW211501002, PW211501003, PW211501004, PW211501010, PW211501011, PW211501012, PW211506001, PW211506002, PW211506003, PW211506004, PW211506005, PW211506006, PW211506007, PW211507001, PW211507002, PW211507003, PW211507004, PW211508001, PW211508002, PW211508004, PW211509001, PW211510003, PW211511004, PW211512002, PW211512003, PW211601005, PW21160113, PW21160114, PW211602001, PW211602005, PW211602006, PW211603004, PW211603005, PW211603006, PW211604003, PW211604004, PW211604007, PW211604008, PW211605008, PW211606001, PW211606002, PW211606003, PW211606004, PW211606006, PW211606008, PW211607002, PW211607003, PW211607006, PW211607007, PW211607010, PW211607011, PW211607012, PW211607013, PW211607014, PW211607015, PW211607016, PW211608002, PW211608003, PW211608004, PW211608005, PW211610003, PW211610004, PW211610005, PW211610009, PW211610010, PW211611006, PW211612001, PW211612002, PW211612005, PW211612009, PW211612012, PW211701002, PW211701003, PW211701004, PW211701005, PW211701006, PW211701007, PW211701008, PW211701011, PW211701012, PW211701014, PW211702001, PW211702002, PW211702003, PW211702004, PW211702005, PW211702006, PW211702007, PW211702008, PW211703001, PW211703002, PW211703003, PW211703005, PW211703006, PW211703007, PW211703008, PW211703009, PW211703010, PW211703011, PW211706001, PW211706002, PW211706003, PW211706004, PW211706005, PW211706006, PW211706007, PW211706010, PW211706011, PW211706012, PW211706013, PW211707001, PW211707002, PW211707003, PW211707004, PW211707010, PW211708001, PW211708002, PW211708003, PW211708004, PW211708005, PW211708006, PW211708007, PW211709001, PW211709002, PW211709003, PW211709005, PW211709006, PW211709007, PW211709009, PW211709010, PW211709012, PW211710002, PW211710003, PW211710004, PW211710005, PW211710006, PW211710007, PW211710008, PW211710011, PW211711003, PW211712002, PW211712003, PW211712004, PW211712005, PW211712006, PW211712007, PW211712008, PW211712009, PW211712010, PW211801001, PW211801002, PW211801005, PW211801006, PW211801007, PW211801008, PW211801009, PW211801010, PW211801011, PW211801012, PW211802001, PW211802002, PW211802003, PW211802004, PW211802005, PW211802008, PW211802009, PW211802012, PW211802014, PW211802015, PW211802016, PW211803001, PW211803002, PW211803003, PW211803004, PW211803005, PW211804001, PW211804002, PW211804003, PW211804004, PW211804010, PW211804011, PW211804012, PW211804013, PW211804014, PW211804015, PW211804016, PW211804017, PW211804018, PW211804019, PW211805001, PW211805004, PW211805005, PW211805006, PW211805007, PW211806007, and PW212612004.

Distribution pattern

Distribution was made to AL, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, MI, NC, ND, NJ, NY, OH, OR, PA, SD, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Argentina, Australia, Canada, China, Czech Republic, Egypt, France, Germany, India, Italy, Republic of Korea, Kuwait, Malaysia, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey, UAE, United Kingdom, and Viet Nam.