Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 84989

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 30, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
U&I CORP.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 - Product Usage:The device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disk disease at one level from C2-C3 disc to the C7-T1 disc. Each product is composed of four components: PEEK cage, titanium plate, and two titanium markers.

Z-1823-2020
Recall number
Z-1823-2020
Initiated
January 30, 2019
Classification
Class II
Status
Terminated
Recalling firm
U&I CORP.
Quantity
42

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less bone-screw engagement than intended by physicians.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less bone-screw engagement than intended by physicians.

Code information

Lots: A18K220, A18L065, A18L013, A18L014, A18L015, A18L018, A18L019, A18L020, A18L021

Distribution pattern

US Nationwide distribution in the state of CA.