device · product 1 of 1
Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen; Software version 02.00.00.0076 - Product Usage: The System is intended to be used as an adjunct to a moderate intensity diet and behavior modification program. All balloons must be removed 6 months after the first balloon is placed. The Obalon Balloons are administered using the Navigation System.
- Recall number
- Z-1805-2020
- Initiated
- May 23, 2019
- Classification
- Class II
- Status
- Completed
- Recalling firm
- Obalon Therapeutics Inc
- Quantity
- 45 units
App-derived interpretation
During initial set-up and/or replacement of the dispenser batteries, the touch dispenser touchscreen can inadvertently lose calibration. This issue can also occur if the touchscreen is pressed when the device is powered on in preparation for a balloon administration. As a result, the touchscreen can become unresponsive and balloon inflation cannot be initiated.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
During initial set-up and/or replacement of the dispenser batteries, the touch dispenser touchscreen can inadvertently lose calibration. This issue can also occur if the touchscreen is pressed when the device is powered on in preparation for a balloon administration. As a result, the touchscreen can become unresponsive and balloon inflation cannot be initiated.
Code information
ObalonR Balloon System with Navigation and Touch. UDI: 00859810006104; Model No.: Model 4300 Touch Dispenser;Software version 02.00.00.0076. Serial Numbers: 0097, 0076, 0077, 0078, 0079, 0080, 0082, 0083, 0084, 0085, 0086, 0087 0088, 0089, 0090, 0091, 0092, 0093, 0094, 0095, 0097, 0099, 0100, 0101, 0102, 0103, 0104, 0106, 0107, 0109, 0111, 0112, 0113, 0115, 0118, 0119, 0120, 0121, 0122, 0125, 0126, 0127, 0129, 0130, 0133.
Distribution pattern
US Nationwide distribution including the states of VA, NY, GA, TX, TN, OR, FL, LA, CT, and NJ.