natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph
- Recall number
- Z-1630-2020
- Initiated
- March 05, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
- Quantity
- 40 units
App-derived interpretation
The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.
Code information
Serial Numbers: OBM00002G2119, OBM00002G2130, OBM00002G2131, OBM00002G2132, OBM00002G2133, OBM00002G2134, OBM00002G2135, OBM00002G2136, OBM00002G2137, OBM00002G2138, OBM00002G2139, OBM00002G2140, OBM00002G2141, OBM00002G2142, OBM00002G2143, OBM00002G2144, OBM00002G2145, OBM00002G2146, OBM00002G2147, OBM00002G2148, OBM00002G2149, OBM00002G2150, OBM00002G2151, OBM00002G2152, OBM00002G2153, OBM00002G2154, OBM00002G2155, OBM00002G2156, OBM00002G2157, OBM00002G2158, OBM00002G2159, OBM00002G2160, OBM00002G2161, OBM00002G2162, OBM00002G2163, OBM00002G2164, OBM00002G2165, OBM00002G2166, OBM00002G2167, OBM00002G2168
Distribution pattern
Worldwide distribution. US nationwide, Australia, Dominican Republic, United Arab Emirates, Israel, Italy, Hong Kong, Korea, Hungary, Switzerland, Colombia, and United Kingdom