device · product 1 of 2
Synchron Systems Opiate 2000 ng Reagent (OP2 1 x 250); catalogue no. 475006 for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators. The assay provides a rapid screening procedure for presence of the analyte in urine.
- Recall number
- Z-1751-2020
- Initiated
- September 04, 2019
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 1610 units
App-derived interpretation
Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.
Official device-enrichment evidence · Sourced
Material/Component Contamination
Inspect official wording and provenance
Reason for recall
Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.
Code information
Catalog Number 475006 ; UDI: 15099590222031; Lot Numbers: M904147 and older including M810039, M901087 and M904147) and 4 lots OP2 (M805122, M810039, and M901087. .
Distribution pattern
Distributed throughout the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.