PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT
- Recall number
- Z-1796-2020
- Initiated
- March 11, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 1932
App-derived interpretation
During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
Official device-enrichment evidence · Sourced
Other
Inspect official wording and provenance
Reason for recall
During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
Code information
SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659)
Distribution pattern
U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy