Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 85325

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 30, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Spacelabs Healthcare, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.

Z-2261-2020
Recall number
Z-2261-2020
Initiated
August 30, 2019
Classification
Class II
Status
Terminated
Quantity
2197 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Manufacturing/Software Deployment

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

Code information

Model No. 96102 Xhibit Central; Software Versions: 1.3.1, 1.3.2, 1.3.3 and 1.3.4; UDI: 10841522100345

Distribution pattern