device · product 1 of 2
ABL90 FLEX PLUS, model no. 393-02; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
These labels are deterministic app interpretations, not FDA categories.
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
Code information
ABL90 FLEX PLUS Model no. 393-092: R0014 - R0195; and Spare Part Hemolyzer Units (902-980): R0526 - R0941; Model Number 393-092.
Distribution pattern
Worldwide distribution - US Nationwide. Also distributed OUS.
device · product 2 of 2
ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
These labels are deterministic app interpretations, not FDA categories.
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
Code information
ABL90 FLEX Model no. 393-090: R0659 - R0977 and Spare Part Hemolyzer Units (902-980): R0526 - R0941
Distribution pattern
Worldwide distribution - US Nationwide. Also distributed OUS.