Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 85392

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 23, 2019
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Tomtec Imaging Systems Gmbh

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

TOMTEC-ARENA TTA2;software version no. 2.20 and lower; Picture archiving and communications system. Software Versions and applicable UDI: TTA 1.2 release prior to UDI requirement TTA2.00.01 release prior to UDI requirement TTA2.11.01 release prior to UDI requirement TTA2.11.02 release prior to UDI requirement TTA2.20.01 UDI: (01)04260361440143(11)160429(10)20.01 TTA2.20.10 UDI:(01)04260361440143(11)160809(10)20.10 TTA2.20.13 UDI:(01)04260361440143(11)160922(10)20.13 TTA2.20.14 UDI:(01)04260361440143(11)161028(10)20.14 TTA2.20.15 UDI: (01)04260361440143[11]161207[10]20.15 TTA2.20.16 UDI:(01)04260361440143(11)170127(10)20.16 TTA2.20.18 UDI:(01)04260361440143(11)170622(10)20.18

Z-2496-2020
Recall number
Z-2496-2020
Initiated
September 23, 2019
Classification
Class II
Status
Ongoing
Quantity
62 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issue

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm discovered a software issue associated with the Image-Com module/clinical application package (CAP) interface for 3D application in the ARENA software, where measurements from one study might be stored to another study.

Code information

The affected product versions are TOMTEC-ARENA, TTA 2.20 and lower, with modules 4D LV-Analysis, 4D RV-Function, 4D MV-Assessment, 3D Cardio-View, 4D Sono-Scan, Echo-Com, 2D Cardiac Performance Analysis, and QAngio.

Distribution pattern