Event 85441

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 18, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Viewray, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ViewRay System: Model No. 10000 and 20000 for radiation treatment. PN11440 MRIdian System Treatment Planning and Delivery System (TPDS) software

Z-1803-2020
Recall number
Z-1803-2020
Initiated
November 18, 2019
Classification
Class II
Status
Terminated
Recalling firm
Viewray, Inc.
Quantity
32

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software shows this misalignment to the user in the predicted dose-volume histogram (DVH). The misalignment impact may

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Registering dose and structures in the treatment delivery workflow could result in an alignment discrepancy between the imported previously delivered dose and the displayed patient anatomy and structures. The MRIdian TPDS software shows this misalignment to the user in the predicted dose-volume histogram (DVH). The misalignment impact may not be obvious to the user. During plan re-optimization the previously delivered dose would be incorrectly accounted for by the software. This may result in unexpectedly higher or lower dose than the intended dose calculated for the subsequent treatment plan. When this occurs in the treatment planning workflow, the registration fails to align the dose with the image and the user is unable to proceed.

Code information

Model No.10000 - Serial No's:. 100,104,105,112,113 Model No. 20000 - Serial No's: 101,102,107,108,200,201,203,204,206,207,209,211, 213,214,215,216,217,218,221,223,225,227,228,232,238,240,243, 244,247,250

Distribution pattern

Worldwide distribution. US states of MO, WI, CA, FL, MI, NY, MI, IL, and MA. South Korea, Netherlands, Italy, Japan, Germany, UAE, Turkey, France, Israel, Denmark, Switzerland, and United Kingdom. The distributors are assigned to applicable regions. Firm send the Urgent Device Corrections to them and they forwarded them to the sites.