Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 85513

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 06, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips North America, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Philips Brilliance iCT Computed Tomography X-Ray System; Model # 728306; UDI #s (01)00884838059474(21)100783, (01)00884838059474(21)100848

Z-1921-2020
Recall number
Z-1921-2020
Initiated
April 06, 2020
Classification
Class II
Status
Terminated
Quantity
19

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.

Code information

Serial #s 100193, 100408, 100457, 100783, 100356, 100113, 100380, 100318, 100848, 100112, 100742, 100447, 100685, 100375, 100449, 100659, 100158, 100398 Additional serial number as of 5/22/20: 100117

Distribution pattern

Worldwide distribution including one unit distributed to Iowa.

device · product 2 of 3

Philips Brilliance iCT SP Computed Tomography X-Ray System; Model # 728311

Z-1922-2020
Recall number
Z-1922-2020
Initiated
April 06, 2020
Classification
Class II
Status
Terminated
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.

Code information

Serial # 200044 Additional unit as off 5/22/20: 200026

Distribution pattern

Worldwide distribution including one unit distributed to Iowa.

device · product 3 of 3

Philips Brilliance IQon Spectral CT, Computed Tomography X-Ray System; Model # 728332; UDI # (01)00884838059542(21)860067

Z-1923-2020
Recall number
Z-1923-2020
Initiated
April 06, 2020
Classification
Class II
Status
Terminated
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.

Code information

Serial #s 860067, 860023

Distribution pattern

Worldwide distribution including one unit distributed to Iowa.