Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 85553

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 20, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Siemens Cios Alpha VA20-mobile X-Ray system Material # 10308191 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.

Z-1957-2020
Recall number
Z-1957-2020
Initiated
April 20, 2020
Classification
Class II
Status
Terminated
Quantity
4 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector

Code information

Serial Numbers: 12425, 109, 10063, 12946

Distribution pattern

US Nationwide distributions.

device · product 2 of 3

Siemens Cios Alpha VA30-mobile X-Ray system, Material # 11105200 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.

Z-1958-2020
Recall number
Z-1958-2020
Initiated
April 20, 2020
Classification
Class II
Status
Terminated
Quantity
16 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector

Code information

Serial Numbers: 40057 40047 40048 40107 40108 40111 40120 40126 40134 40135 40032 40104 40075 40102 40124 40070

Distribution pattern

US Nationwide distributions.

device · product 3 of 3

Siemens Cios Spin VA30-mobile X-Ray system, Material # 10308194 - Product Usage: The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.

Z-1959-2020
Recall number
Z-1959-2020
Initiated
April 20, 2020
Classification
Class II
Status
Terminated
Quantity
46 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector

Code information

Serial Numbers: 50124 50135 50131 50171 50129 50032 50033 50048 50050 50051 50073 50075 50081 50088 50098 50107 50109 50111 50120 50196 50102 50180 50136 50185 50096 50155 50091 50130 50127 50077 50078 50134 50122 50101 50092 50094 50166 50049 50192 50161 50125 50128 50132 50079 50160 50030

Distribution pattern

US Nationwide distributions.