Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 85652

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 11, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ortho-Clinical Diagnostics

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI # 10758750008513 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Z-2364-2020
Recall number
Z-2364-2020
Initiated
May 11, 2020
Classification
Class II
Status
Terminated
Quantity
73,071

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

Code information

Lot #s 600203564949, 600203564950, 600203564975, 600203564976, 600203564977, 600203564978, 600203566275, 600203566276, 600203566277, 600203566537, 600203566538, 600203566620, 600203566621, 600203566622, 600203566698, 600203566699, 600203566700, 600203567485, 600203567486, 600203567546, 600203567547, 600303570416, 600303570417, 600303570476, 600303570495, 600303570496, 600303571564, 600303571565, 600303576365, 600303576366, 600303578980, 600303578981, 600303578982, 600303578984, 600303578985, 600303579304, 600303579305, 600303579338, 600303579339, 600303579545, 600303579546, 600303579570, 600303579571, 600503581579, 620503581598, 600503581599, 600503581600, 600503581626, 600503582485, 600503582661, 600503582711, 600503582712, 600503582713, 600503583174, 600503583175, 600503584118, 600503584168, 600503584169, 600503589810, 600503589811, 600503589812, 600603591744, 600603591745, 600603591746, 600603595598, 600603595599, 600603595644, 600603595645, 600603595655, 600603595656, 600603597378, 600603597379, 600603599568, 600603599572, 600603599630, 600803637422, 600803637423, 600803637424, 600803638469, 600803638470

Distribution pattern

Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska

device · product 2 of 4

VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/90 SLDS, Product Code 8112724, UDI # 10758750004393 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Z-2365-2020
Recall number
Z-2365-2020
Initiated
May 11, 2020
Classification
Class II
Status
Terminated
Quantity
64,697

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

Code information

Lot #s 600203565250, 600203565563, 600203565573, 600203565902, 600203566266, 600203566278, 600203566676, 600203567387, 600203567388, 600203567389, 600303570323, 600303570338, 600303570824, 600303570830, 600303570847, 600303570848, 600303577945, 600303578593, 600303579040, 600303579041, 600303579565, 600303579566, 600303579616, 600503581489, 600503581495, 600503582261, 600503582262, 600503582269, 600503582570, 600503582607, 600503583117, 600503584003, 600503584704, 600503584705, 600503584706, 600503584987, 600503585015, 600503585951, 600503585957, 600503586204, 600503586207, 600503586219, 600603594331, 600603596567, 600603596601, 600603596692, 600603596693, 600803638425, 600803638426, 600803638483, 600803638489, 600803638490, 600803639060, 600803639078, 600803639091

Distribution pattern

Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska

device · product 3 of 4

VITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Z-2366-2020
Recall number
Z-2366-2020
Initiated
May 11, 2020
Classification
Class II
Status
Terminated
Quantity
Unavailable

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

Code information

Current Control Assay Sheet SD 4.2, 4.3 Revised Control Assay Sheet SD 6.8, 7.1

Distribution pattern

Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska

device · product 4 of 4

VITROS Chemistry Product Performance Verifier II, Product Code 8067324, UDI # 10758750004317 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Z-2367-2020
Recall number
Z-2367-2020
Initiated
May 11, 2020
Classification
Class II
Status
Terminated
Quantity
Unavailable

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

Code information

Current Control Assay Sheet SD 10.5, 10.8 Revised Control Assay Sheet 17.1, 17.6

Distribution pattern

Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska