Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 85822

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 10, 2020
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Intuitive Surgical, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.

Z-2949-2020
Recall number
Z-2949-2020
Initiated
June 10, 2020
Classification
Class II
Status
Ongoing
Recalling firm
Intuitive Surgical, Inc.
Quantity
67 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.

Code information

All lots.

Distribution pattern

U.S.: California, Colorado, Florida, Illinois, Massachusetts, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, and Virginia. No foreign consignees.

device · product 2 of 2

ION Reprocessing Accessories Kit, IF1000 (Part Number: 490024-03) UDI:00886874119129. The kit includes an ION Catheter Reprocessing Cover (Part Number: 490114-02).

Z-2950-2020
Recall number
Z-2950-2020
Initiated
June 10, 2020
Classification
Class II
Status
Ongoing
Recalling firm
Intuitive Surgical, Inc.
Quantity
27 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.

Code information

All lots.

Distribution pattern

U.S.: California, Colorado, Florida, Illinois, Massachusetts, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, and Virginia. No foreign consignees.