openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 2
Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
These labels are deterministic app interpretations, not FDA categories.
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
Code information
All lots.
Distribution pattern
U.S.: California, Colorado, Florida, Illinois, Massachusetts, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, and Virginia. No foreign consignees.
device · product 2 of 2
ION Reprocessing Accessories Kit, IF1000 (Part Number: 490024-03) UDI:00886874119129. The kit includes an ION Catheter Reprocessing Cover (Part Number: 490114-02).
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
These labels are deterministic app interpretations, not FDA categories.
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
Code information
All lots.
Distribution pattern
U.S.: California, Colorado, Florida, Illinois, Massachusetts, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, and Virginia. No foreign consignees.