openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is intended for pipetting samples into a MagNA Pure 96 Processing Cartridge (purification rack position 1) before a purification run.
When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.
These labels are deterministic app interpretations, not FDA categories.
When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.
Code information
All units when used with Sample Transfer protocol version 3.0
Distribution pattern
Worldwide distribution - US Nationwide distribution.