Software 1898072 IPC upgrade v 2.7.3.0 - Product Usage: The IPC system is indicated for the incision/cutting removal, drilling and sawing of soft and hard tissue and bone in surgical procedures.
- Recall number
- Z-2673-2020
- Initiated
- June 12, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Medtronic Xomed, Inc.
- Quantity
- 673 units
App-derived interpretation
During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.
Code information
Model: 1898072 Affected Lot Numbers: 0217355465, 0218377490, 0218687535, 0218928638, 0219310085,0219639840, 0217856852, 0218472653, 0218714701, 0218958034, 0219443322,0219639842, 0217858947,0218472654, 0218721203, 0218958035, 0219448447, 0219639844, 0217859045,0218665536, 0218745073, 0219004569,0219465587, 0218203285, 0218666231, 0218746389,0219201122,0219518520,0218263964, 0218666243, 0218864105, 0219255151, 0219550550, 0218293282, 0218687534, 0218900337,0219281168, 0219619754.
Distribution pattern
Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA,CO, CT, FL, GA, HI, ID,IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OK, PA, RI, SC, TN, TX, UT, VA, WV, WA, WI. The countries of Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, United Kingdom.