Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 85889

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 10, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
NuVasive Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Nuvasive NVM5 EMG Module, REF 8020015, Rx ONLY, CE2797, Nonsterile, UDI: 00887517206633 - Product Usage: is intended for intraoperative neurophysiologic monitoring during spinal surgery.

Z-2684-2020
Recall number
Z-2684-2020
Initiated
June 10, 2020
Classification
Class II
Status
Terminated
Recalling firm
NuVasive Inc
Quantity
23

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The NVM5 EMG Module kit intended for intraoperative neurophysiologic monitoring during spinal surgery contains incorrect component, part number 1715945, SSEP harness instead of the the correct part number 1747065, NVM5 Harness, Gen 3.5 EM. This issue can result in delay in therapy or lack of EMG neuromonitoring during surgery if the surgeon elects to proceed without the NVM5.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The NVM5 EMG Module kit intended for intraoperative neurophysiologic monitoring during spinal surgery contains incorrect component, part number 1715945, SSEP harness instead of the the correct part number 1747065, NVM5 Harness, Gen 3.5 EM. This issue can result in delay in therapy or lack of EMG neuromonitoring during surgery if the surgeon elects to proceed without the NVM5.

Code information

Lot 1075097

Distribution pattern

US Nationwide distribution including in the states of AR, CA, IA, IL, WI. OUS: None