Alaris System PC Unit Model 8015
- Recall number
- Z-2909-2020
- Initiated
- August 04, 2020
- Classification
- Class I
- Status
- Ongoing
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 313740
App-derived interpretation
The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.
Official device-enrichment evidence · Sourced
Reprocessing Controls
Inspect official wording and provenance
Reason for recall
The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.
Code information
All serial numbers manufactured from April 07, 2017 to Present.
Distribution pattern