Alaris Pump Module Model 8100
- Recall number
- Z-2736-2020
- Initiated
- June 30, 2020
- Classification
- Class I
- Status
- Terminated
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 1093588
App-derived interpretation
Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame assembly. These may prevent devices from delivering accurate amounts of fluid, which may result in an over infusion, free-flow conditions, or under infusion.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame assembly. These may prevent devices from delivering accurate amounts of fluid, which may result in an over infusion, free-flow conditions, or under infusion.
Code information
All serial numbers
Distribution pattern
US: CA, FL, IN, TX, NC, LA, SC, VA, PA, OK, AZ, IL, KY, UT, AL, GA, TN, MA, MI, OH, AK, OR, MS, CO, NY, WI, KS, MO, MD, NV, AR, AS, ID, VT, NE, NJ, DC, HI, CT, MT, WV, NH, WA, DE, ME, NM, MN, WY, IA, SD, ND, MP, PR, GU. OUS: NL, MX, AU, MY, NZ, ZA, GB, FR, CH, CN, SG, TW, CA, IL, KW, QA, DE, BH, IN, SA, GI, AE, PK, PH, BE