Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 86000

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 24, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

syngo.via RT Image Suite with software versions syngo.via VB30 or VB40

Z-2688-2020
Recall number
Z-2688-2020
Initiated
June 24, 2020
Classification
Class II
Status
Terminated
Quantity
21 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software does not recognize the adapted acquisition plane. This may result in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the software does not recognize the adapted acquisition plane. This may result in images with wrong geometry. When this distortion remains unnoticed and the images are subsequently exported to a treatment planning system (TPS), an incorrect calculated radiation treatment plan cannot be excluded. The occurence of this issue is very unlikely and has never been reported so far.

Code information

syngo.via VB30 or VB40: All Lots Model #: 10496180

Distribution pattern

US Nationwide distribution including in the states of AZ, MA, WI, FL, PA. TX, MD, LA, TX, MN, MI, NY, SC, OH, GA, CA, IL, MO, CA, TX.