device · product 1 of 1
Holder HKHZ 19 for use with Quadrox-i and Quadrox-iD Neonatal / Pediatric Oxygenators, Model Number: 701047495 - Product Usage: The Holder HKHZ 19 is an accessory to QUADROX i and QUADROX iD Neonatal / Pediatric oxygenator and is being used as a single holder for QUADROX i and QUADROX iD Neonatal / Pediatric Oxygenator and can also be used in combination with the holder's arm (016061) to use in cardiopulmonary bypass procedures (CPB) in a pediatric patient population.
- Recall number
- Z-2780-2020
- Initiated
- June 18, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Quantity
- 50 units
App-derived interpretation
Removal for a specific lot, lot number 70128207, of Holder HKHZ 19 for use with QUADROX iD Pediatric Oxygenator. The Holder may not always snap and securely hold the oxygenator as intended.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Removal for a specific lot, lot number 70128207, of Holder HKHZ 19 for use with QUADROX iD Pediatric Oxygenator. The Holder may not always snap and securely hold the oxygenator as intended.
Code information
Model Number: 701047495 Part Number: HKHZ 19 UDI Code: 04037691533339 Lot# 70128207
Distribution pattern
Worldwide distribution - US Nationwide distribution including in the states of OH, PA, SC, CT and the countries of Germany, Spain, France, United Kingdom, Italy, United Arab Emirates, Mexico, Canada, Croatia, Georgia, Poland, Slovakia, Thailand, and Vietnam.