The Guardian Connect App CSS7200 iOS - Product Usage: intended for use by patients with a compatible consumer mobile device.
- Recall number
- Z-2980-2020
- Initiated
- July 15, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Medtronic Inc.
- Quantity
- 3,588 users
App-derived interpretation
As a result of the release of new software version to CareLink Personal website, the IOS app for the Continuous Monitoring Glucose system stopped automatically uploading data to the website.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
As a result of the release of new software version to CareLink Personal website, the IOS app for the Continuous Monitoring Glucose system stopped automatically uploading data to the website.
Code information
Product Number - CSS7200 UDI (GTIN.UPN) - 00763000255404
Distribution pattern
International distribution including in the countries of EMEA - Austria, Belgium, Croatia, Czechia, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Kuwait, Latvia, Luxembourg, The Netherlands, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, and United Kingdom. Latin America- Chile, Colombia and Mexico Asia Pacific (APAC) - Australia, India, Japan, The Republic of Korea, Malaysia, New Zealand, Russia Federation, Singapore, Thailand, and Vietnam. Asia - Hong Kong North America - Canada