device · product 1 of 1
Panther Fusion Tube Tray, Part: FAB-15004, REF: PRD-04000
- Recall number
- Z-2990-2020
- Initiated
- August 04, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Hologic, Inc
- Quantity
- 607
App-derived interpretation
Tube trays may leak which could invalidate patient results or assay worklists, causing delayed results for any assays with which they are used. A possibility exists of false negative result with flu A/B/RSV assay or paraflu assay.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Tube trays may leak which could invalidate patient results or assay worklists, causing delayed results for any assays with which they are used. A possibility exists of false negative result with flu A/B/RSV assay or paraflu assay.
Code information
Box/ Tray lots: 271436/ 618539, 271438/ 620899, 272089/ 620899, 272959/ 621837, 272970/ 621256, 272971/ 621837, 274670/ 625351
Distribution pattern