Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 86255

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 13, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Clinical Innovations, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Clinical Innovations ROM Plus Collection Kit IFU: ROM Plus Rupture of Membrane Test ROM-5025, ROM-6025

Z-0541-2021
Recall number
Z-0541-2021
Initiated
August 13, 2020
Classification
Class II
Status
Terminated
Recalling firm
Clinical Innovations, LLC
Quantity
19450 units/kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible false negative or false positive results. The incorrect sample collection swabs (non-scored) were assembled into assembled into the in vitro diagnostic kits. If the user does not break the swab stick correctly, or if scissors or other tools are used to cut the swab handle false negative or false positive results can occur.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible false negative or false positive results. The incorrect sample collection swabs (non-scored) were assembled into assembled into the in vitro diagnostic kits. If the user does not break the swab stick correctly, or if scissors or other tools are used to cut the swab handle false negative or false positive results can occur.

Code information

Model Number: ROM-5025 Lot Number: 192438 Model Number: ROM-6025 Lot Numbers: 192353, 192359, 192631

Distribution pattern

US: CO, IL, IN, MO, OH, RI, UT, and VA. OUS: None