Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 86364

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 12, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Zeus Scientific, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ZEUS ELISA Parvovirus B19 IgM Test System, UDI 00845533001755 - Product Usage: intended for the qualitative detection of IgM class antibodies to human Parvovirus B19 in human serum including women of childbearing age where there is a suspicion of exposure to human Parvovirus B19.

Z-3026-2020
Recall number
Z-3026-2020
Initiated
August 12, 2019
Classification
Class II
Status
Terminated
Recalling firm
Zeus Scientific, Inc.
Quantity
74

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.

Code information

Lot 19040014

Distribution pattern

Worldwide distribution - US Nationwide distribution including in the states of CA, CT, IN, NJ, TX and the country of Canada.