Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 86476

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 16, 2020
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Boston Scientific Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Hurricane RX Biliary Balloon Dilatation Catheter Catalog REF No. M00545890; M00545900; M00545910; M00545920; M00545930; M00545940; M00545950; M00545960 The device is a double lumen catheter with a balloon mounted at the distal tip. Dilatation balloon catheters are used to exert radial force to dilate narrow duct segments, as well as the Sphincter of Oddi.

Z-0286-2021
Recall number
Z-0286-2021
Initiated
September 16, 2020
Classification
Class II
Status
Ongoing
Quantity
15,062 devices distributed in US. 12,930 distributed OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RX tunnel component (black sheath) may detach from the catheter shaft due to lower bond strength.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Vendor change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RX tunnel component (black sheath) may detach from the catheter shaft due to lower bond strength.

Code information

Hurricane RX Dilation Balloon 4 mm 2 cm UPN/REF # M00545890 GTIN # 08714729283805 Lot/Batch # 25509445, 25536219, 25612668, 25649812, 25704208 EXP Date Range 04/13/2022 - 06/13/2022 Hurricane RX Dilation Balloon 4 mm 4 cm UPN/REF # M00545900 GTIN # 08714729283812 Lot/Batch # 25326198, 25344165, 25344168, 25366542, 25366543, 25379240, 25379241, 25392546, 25447687, 25447688, 25447689, 25447690, 25447691, 25447692, 25453103, 25453105, 25453108, 25466435, 25468596, 25468597, 25471588, 25479171, 25479176, 25479177, 25479178, 25479521, 25481983, 25481984, 25481985, 25481986, 25497466, 25497467, 25497468, 25506678, 25509440, 25509446, 25509447, 25509450, 25518274, 25518305, 25518314, 25533944, 25536416, 25536680, 25536860, 25543967, 25543969, 25571363, 25575934, 25575935, 25576717, 25576718, 25576719, 25576720, 25576721, 25576769, 25611564, 25612669, 25612670, 25612671, 25612673, 25612674, 25623377, 25623378, 25623379, 25623380, 25623381, 25623382, 25623383, 25623384, 25640115, 25649813, 25649816, 25678821, 25678826, 25678831, 25678836, 25680096, 25704209, 25704210, 25704211, 25704212, 25704213, 25704214, 25704215, 25704216, 25739146, 25739147, 25739149, 25739150, 25771761, 25771762, 25771765 EXP Date Range 02/07/2022 - 06/27/2022 Hurricane RX Dilation Balloon 6 mm 2 cm UPN/REF # M00545910 GTIN #08714729283829 Lot/Batch # 25456946, 25497472, 25571369, 25623385, 25678822, 25739151 EXP Date Range 03/22/2022 - 06/20/2022 Hurricane RX Dilation Balloon 6 mm 4 cm UPN/REF # M00545920 GTIN # 08714729283836 Lot/Batch # 25305803, 25326193, 25326194, 25326195, 25326199, 25334817, 25334818, 25366544, 25366545, 25366546, 25366547, 25392549, 25392551, 25392553, 25392554, 25392555, 25397943, 25397944, 25447697, 25447698, 25453106, 25453109, 25461680, 25466436, 25468598, 25472425, 25475235, 25475236, 25479172, 25479173, 25479179, 25481987, 25481988, 25481989, 25481990, 25481991, 25497469, 25497470, 25497473, 25497474, 25497475, 25509412, 25509413, 25509414, 25509451, 25509452, 25509453, 25518276, 25518306, 25518307, 25518315, 25518316, 25536681, 25536739, 25536740, 25543970, 25543971, 25543972, 25543973, 25571370, 25571373, 25576770, 25576771, 25576772, 25576773, 25576881, 25576882, 25576883, 25576884, 25576885, 25608762, 25623386, 25623387, 25623388, 25623389, 25623390, 25623391, 25623392, 25623393, 25623394, 25623395, 25623396, 25623397, 25623398, 25649814, 25649817, 25649820, 25649821, 25649825, 25649826, 25649830, 25649831, 25678823, 25678827, 25678832, 25678837, 25680097, 25704217, 25704218, 25704219, 25704220, 25704221, 25704222, 25704223, 25739152, 25739153, 25739154, 25739155, 25739157, 25739158, 25771773 EXP Date Range 02/02/2022 - 06/27/2022 Hurricane RX Dilation Balloon 8 mm 2 cm UPN/REF # M00545930 GTIN # 08714729283843 Lot/Batch # 25509441, 25536741, 25644951, 25649190, 25650286, 25651334, 25654287 EXP Date Range 04/13/2022 - 05/31/2022 Hurricane RX Dilation Balloon 8 mm 4 cm UPN/REF # M00545940 GTIN # 08714729283850 Lot/Batch # 25317077, 25326196, 25326220, 25326221, 25326222, 25334819, 25344169, 25366548, 25366549, 25392556, 25392557, 25397079, 25447693, 25447694, 25453104, 25453107, 25453110, 25466437, 25468599, 25471589, 25471590, 25475234, 25479174, 25479175, 25479520, 25479580, 25481992, 25497051, 25497052, 25502738, 25509415, 25509416, 25509417, 25509418, 25509419, 25509442, 25509448, 25509454, 25509546, 25518277, 25518278, 25518317, 25518318, 25536682, 25536742, 25536743, 25543975, 25543976, 25543977, 25543978, 25543979, 25544020, 25571371, 25571374, 25596599, 25602458, 25608760, 25612672, 25649815, 25649818, 25649822, 25649823, 25649827, 25649828, 25649832, 25649833, 25678824, 25678828, 25678829, 25678833, 25678834, 25678838, 25680098, 25680099, 25704224, 25704225, 25704226, 25704227, 25704228, 25704229, 25739159, 25739160, 25739161, 25739162, 25739163, 25739164, 25739165, 25739166, 25739167, 25739168, 25788242, 25789369, 25789801 EXP Date Range 02/04/2022 - 06/30/2022 Hurricane Rx Dilation Balloon 10 mm 2 cm UPN/REF # M00545950 GTIN # 08714729283867 Lot/Batch # 25471587, 25571372, 25588303, 25673064 EXP Date Range 03/28/2022 - 06/06/2022 Hurricane RX Dilation Balloon 10 mm 4 cm UPN/REF # M00545960 GTIN # 08714729283874 Lot/Batch # 25334980, 25344170, 25344171, 25366550, 25378999, 25397945, 25397950, 25447685, 25447695, 25447696, 25461681, 25464817, 25466438, 25471591, 25471592, 25484487, 25493901, 25497471, 25497476, 25509443, 25509444, 25509449, 25518279, 25518308, 25518309, 25536683, 25544022, 25544023, 25544024, 25544025, 25571375, 25602323, 25608761, 25640116, 25640117, 25644834, 25649819, 25649824, 25649829, 25649834, 25678825, 25678830, 25678835, 25678839, 25678880, 25680560, 25704230, 25704231, 25704232, 25739169, 25739170 EXP Date 02/08/2022 - 06/20/2022

Distribution pattern

Worldwide distribution - US Nationwide and International distribution.