Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 86534

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 22, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Boston Scientific Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application

Z-0319-2021
Recall number
Z-0319-2021
Initiated
September 22, 2020
Classification
Class II
Status
Terminated
Quantity
8 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.

Code information

Model 3892 v1.04 is installed on Model 3300 LATITUDE Programming Systems with the following serial numbers: 009494 012689 010356 008206 008124 007983 004813 010751

Distribution pattern

Distributed in US - AR, FL, IL, and TX