Event 86681

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 23, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medtronic Heart Valves Division

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Evolut (TM) R System. With the following model numbers: a. EVOLUTR-23, Labeled as VLV EVOLUTR-23 VT IR 6L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-23 TAV CE AU 2Y MX, VLV EVOLUTR-23 TAV JPN MX, VLV EVOLUTR-23 GREY 29L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-23 TAV EVOLUT R CE 26L MX, VLV EVOLUTR-23 TAV CA 2Y MX, VLV EVOLUTR-23 GREY 6L COMM OUS, VLV EVOLUTR-23-US TAV 23 US COMM MX; b. EVOLUTR-23-C, Labeled as VLV EVOLUTR-23-C TAV EVOLUT R US IDE MEX; c. EVOLUTR-23-US, Labeled as VLV EVOLUTR-23-US GREY COMM US, VLV EVOLUTR-23-US TAV 23 US COMM MX; d. EVOLUTR-26, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX VLV EVOLUTR-26 VT IR 6L COMM OUS, VLV EVOLUTR-26 TAV EVOLUT R CE 7L MX, LV EVOLUTR-26 TAV CE AU 2Y MX, VLV EVOLUTR-26 TAV JPN MX, VLV EVOLUTR-26 GREY 29L COMM OUS, VLV EVOLUTR-26-US TAV 26 US COMM MX, VLV EVOLUTR-26 TAV CLN OUS MX, VLV EVOLUTR-26 TAV CA 2Y MX, VLV EVOLUTR-26 TAV CE CA 2Y MX, VLV EVOLUTR-26 GREY 6L COMM OUS; e. EVOLUTR-26-C, Labeled as VLV EVOLUTR-26-C TAV EVOLUT R US IDE MEX, VLV EVOLUTR-26-C TAV CLN US MX; f. EVOLUTR-26-US, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-26-US GREY COMM US, VLV EVOLUTR-26-US TAV 26 US COMM MX; g. EVOLUTR-29, Labeled as VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-29 GREY 6L COMM OUS, VLV EVOLUTR-29 VT IR 6L COMM OUS, VLV EVOLUTR-29 TAV CE AU 2Y MX, VLV EVOLUTR-29 TAV JPN MX, VLV EVOLUTR-29 GREY 29L COMM OUS, VLV EVOLUTR-29 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-29 TAV CLN OUS MX, VLV EVOLUTR-29 TAV CA 2Y MX, VLV EVOLUTR-29 TAV CE CA 2Y MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; h. EVOLUTR-29-C, Labeled as VLV EVOLUTR-29-C TAV EVOLUT R US IDE MEX, LV EVOLUTR-29-C TAV CLN US MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; i. EVOLUTR-29-US, Labeled as VLV EVOLUTR-29-US GREY COMM US, VLV EVOLUTR-29-US TAV 29 US COMM MX, VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX; j. EVOLUTR-34, Labeled as VLV EVOLUTR-34 BLUE 29L COMM OUS, VLV EVOLUTR-34 VT IR 6L COMM OUS, VLV EVOLUTR-34 BLUE 6L COMM OUS, VLV EVOLUTR-34 TAV CE AU 2Y MX, VLV EVOLUTR-34 TAV JPN MX, VLV EVOLUTR-34 TAV CLN OUS MX, VLV EVOLUTR-34 TAV CA 2Y MX, VLV EVOLUTR-34 TAV CE CA 2Y MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34 TAV EVO R CE 29L 2Y MX SA, DRAPE 9732722 TUBE STERILE O-ARM 20PK; k. EVOLUTR-34-C, Labeled as VLV EVOLUTR-34-C TAV EVOLUT R US IDE MX, VLV EVOLUTR-34-C TAV CLN US MX; l. EVOLUTR-34-US, Labeled as VLV EVOLUTR-34-US BLUE COMM US, VLV EVOLUTR-34-US TAV 34 US COMM MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34-US TAV 34 US COMM MX SA.

Z-0559-2021
Recall number
Z-0559-2021
Initiated
October 23, 2020
Classification
Class II
Status
Terminated
Quantity
172,851 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

Code information

All GTINs and associated serial numbers (refer to Attachment A-Product List as of 22-Oct-2020).

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, Norway, Egypt, Saudi Arabia.

device · product 2 of 3

Evolut (TM) PRO System. With the following model numbers: a. EVOLUTPRO-23, Labeled as: VLV EVOLUTPRO-23 PRO COMM 29L OUS, VLV EVOLUTPRO-23 BLUE 29L COMM OUS, VLV EVOLUTPRO-23 BLUE COMM JPN, VLV EVOLUTPRO-23 BLUE 6L COMM OUS, VLV EVOLUTPRO-23 PRO COMM 29L CAN, VLV EVOLUTPRO-23 PRO COMM 29L OUS SA, VLV EVOLUTPRO-23 PRO COMM 29L AUS; b. EVOLUTPRO-23-US, Labeled as: VLV EVOLUTPRO-23-US BLUE COMM US, VLV EVOLUTPRO-23-US PRO BLNK US; c. EVOLUTPRO-26, Labeled as: VLV EVOLUTPRO-26 PRO COMM 29L OUS, VLV EVOLUTPRO-26 BLUE 6L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L AUS, VLV EVOLUTPRO-26 BLUE COMM JPN, VLV EVOLUTPRO-26 BLUE 29L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L CAN, VLV EVOLUTPRO-26 PRO COMM 29L OUS SA; d. EVOLUTPRO-26-US, Labeled as: VLV EVOLUTPRO-26-US BLUE COMM US, VLV EVOLUTPRO-26-US PRO BLNK US; e. EVOLUTPRO-29, Labeled as: VLV EVOLUTPRO-29 PRO COMM 29L OUS, VLV EVOLUTPRO-29 BLUE 29L COMM OUS, VLV EVOLUTPRO-29 BLUE 6L COMM OUS, VLV EVOLUTPRO-29 PRO COMM 29L AUS, VLV EVOLUTPRO-29 BLUE COMM JPN, VLV EVOLUTPRO-29 PRO COMM 29L CAN, VLV EVOLUTPRO-29 PRO COMM 29L OUS SA; f. EVOLUTPRO-29-US, Labeled as: VLV EVOLUTPRO-29-US BLUE COMM US, VLV EVOLUTPRO-29-US PRO BLNK US g. , Labeled as:

Z-0560-2021
Recall number
Z-0560-2021
Initiated
October 23, 2020
Classification
Class II
Status
Terminated
Quantity
71,030 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

Code information

All GTINs and associated serial numbers (refer to Attachment A-Product List as of 22-Oct-2020).

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, Norway, Egypt, Saudi Arabia.

device · product 3 of 3

Evolut (TM) PRO+ System. With the following model numbers: a. EVPROPLUS-23US, Labeled as VLV EVPROPLUS-23US COMM JPN, VLV EVPROPLUS-23 COMM US; b. EVPROPLUS-26US, Labeled as VLV EVPROPLUS-26US COMM JPN, VLV EVPROPLUS-26 COMM US; c. EVPROPLUS-29US, Labeled as VLV EVPROPLUS-29US COMM JPN, VLV EVPROPLUS-29 COMM US; d. EVPROPLUS-34US, Labeled as VLV EVPROPLUS-34 COMM US, VLV EVPROPLUS-34 COMM US SA.

Z-0561-2021
Recall number
Z-0561-2021
Initiated
October 23, 2020
Classification
Class II
Status
Terminated
Quantity
19,961 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

Code information

All GTINs and associated serial numbers (refer to Attachment A-Product List as of 22-Oct-2020).

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, Norway, Egypt, Saudi Arabia.