Event 86713

17 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 19, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Synthes (USA) Products LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

17 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 17

DePuy Synthes 10.0MM REAMER HEAD FOR RIA 2, STERILE-Orthopedic manual surgical instrument Part Code: 03.404.016S

Z-0571-2021
Recall number
Z-0571-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1297 US 528 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982273932

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 2 of 17

DePuy Synthes 10.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.017S

Z-0572-2021
Recall number
Z-0572-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1066 US; 397 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982273949

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 3 of 17

DePuy Synthes 11.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.018S

Z-0573-2021
Recall number
Z-0573-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1275 US and 445 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982273963

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 4 of 17

DePuy Synthes 11.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.019S

Z-0574-2021
Recall number
Z-0574-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1231US; 422 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982273970

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 5 of 17

DePuy Synthes 12.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.020S

Z-0575-2021
Recall number
Z-0575-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1566 US ; 732 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982273987

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 6 of 17

DePuy Synthes 12.5MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.021S

Z-0576-2021
Recall number
Z-0576-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1337 US; 453 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274007

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 7 of 17

DePuy Synthes 13.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.022S

Z-0577-2021
Recall number
Z-0577-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1447 US; 554 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274021

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 8 of 17

DePuy Synthes 13.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.023S

Z-0578-2021
Recall number
Z-0578-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1262 US; 549 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274045

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 9 of 17

DePuy Synthes 14.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.024S

Z-0579-2021
Recall number
Z-0579-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1328 US ; 578 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274069

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 10 of 17

DePuy Synthes 14.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.025S

Z-0580-2021
Recall number
Z-0580-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
1142 US; 450 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274083

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 11 of 17

DePuy Synthes 15.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.026S

Z-0581-2021
Recall number
Z-0581-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
748 US; 381 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982273956

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 12 of 17

DePuy Synthes 15.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.027S

Z-0582-2021
Recall number
Z-0582-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
642 US; 325 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI:10886982273994

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 13 of 17

DePuy Synthes 16.0MM REAMER HEAD FOR RIA 2 STERILE--Orthopedic manual surgical instrument Part Code: 03.404.028S

Z-0583-2021
Recall number
Z-0583-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
651US : 326 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274014

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 14 of 17

DePuy Synthes 16.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.029S

Z-0584-2021
Recall number
Z-0584-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
531 US; 317 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274038

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 15 of 17

DePuy Synthes 17.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.030S

Z-0585-2021
Recall number
Z-0585-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
572 US; 200 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274052

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 16 of 17

DePuy Synthes 17.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.031S

Z-0586-2021
Recall number
Z-0586-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
515 US; 150 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274076

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,

device · product 17 of 17

DePuy Synthes 18.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.032S

Z-0587-2021
Recall number
Z-0587-2021
Initiated
November 19, 2020
Classification
Class II
Status
Terminated
Quantity
542 US ; 154 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed

Code information

All Lots UDI: 10886982274090

Distribution pattern

Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,