Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18S, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/6f Straight Catheter, Sterile EO,
- Recall number
- Z-1072-2021
- Initiated
- December 10, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Argon Medical Devices, Inc
- Quantity
- 483 units
App-derived interpretation
manufacturing error
Official device-enrichment evidence · Sourced
Process change control
Inspect official wording and provenance
Reason for recall
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Code information
Lot Number: 1419196, 1460291, 1469844, 1502325
Distribution pattern
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain