openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
These labels are deterministic app interpretations, not FDA categories.
When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
Code information
Model number: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Pheno- 10849000 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355