Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87055

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 07, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Carl Zeiss Meditec AG

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

IOLMaster 700

Z-1133-2021
Recall number
Z-1133-2021
Initiated
December 07, 2020
Classification
Class II
Status
Terminated
Recalling firm
Carl Zeiss Meditec AG
Quantity
613 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen

Code information

Model: IOLMaster 700 Catalog Number device: 000000-1932-169

Distribution pattern

U.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Austria, Australia, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, France, United Kingdom, Guatemala, Hong Kong, Indonesia, Ireland, India, Italy, Japan, Kuwait, Malaysia, Netherlands, New Zealand, Philippines, Poland, Portugal, Russia, Sweden, Singapore, Slovakia, Thailand, Taiwan, and South Africa.