Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87155

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 21, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Getinge Group Logistics America, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Getinge CM320 Series Washer Disinfector - Product Usage: A multi chamber washer disinfector intended for washing, disinfecting and drying of surgical instruments, containers, operating theatre shoes and operating theatre accessories normally used in hospitals and central disinfection facilities. The disinfected items must be further processed through terminal sterilization before use in any surgical procedure.

Z-1124-2021
Recall number
Z-1124-2021
Initiated
December 21, 2020
Classification
Class II
Status
Terminated
Quantity
189 units total (Domestic 19 units, Foreign 170 units)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.

Code information

Product Code/Part Number: CM320 4 WUWD, CM320 3 WWD, CM320K661, CM320K671, CM320K689, CM320K706, CM320K730, CM320WU, CM320WU K981, CM320WUWD. Serial Number: 31253, 31260, 31299, 31286, W50026718, W50027250, W50029991, W50034570, W50039234, WAA070170, WAA081312, WAA062340, WAA062341, WAA070619, WAA085063, WAA089556, WAA093936, WAA094247, WAA070618. UDI Code: 07340153700178 GETINGE CM320WO CM320 2 CHAMBER WO 07340153700185 GETINGE CM320WU CM320 2 CHAMBER WU 07340153700192 GET/NGE CM320WUWO CM320 4 CHAMBER WUWO 07340153700208 GETINGE CM320WWO CM320 3 CHAMBER WWO 07340153700215 GETINGE CM320WWWO CM320 4 CHAMBER WWWO 07340153710153 GETINGE CM320WOK966 CM320 2 CHAMBER WO

Distribution pattern

Worldwide distribution - US Nationwide distribution including in the states of ND, PA, NC, VA, NE, MD, NY, MN and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Norway, Slovakia, Slovenia, South Korea, Sweden, Japan, China, Netherland, Italy, Kuwait and Czech Republic.