Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87161

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 14, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Phadia US Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

EliA dsDNA Well - Product Usage: intended for the In-Vitro quantitative measurement of IgG antibodies directed to dsDNA in human serum and plasma as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE). Material Number: 14-5500-01

Z-1107-2021
Recall number
Z-1107-2021
Initiated
December 14, 2020
Classification
Class II
Status
Terminated
Recalling firm
Phadia US Inc
Quantity
365 kits US

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).

Code information

Carrier Lot Number: BFA3Y; Kit Lot Number: 0142; (UDI): 07333066019406

Distribution pattern

US Nationwide distribution.