Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87305

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 01, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Insulet Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Omnipod DASH Insulin Management System User Guide (OUS Only) Product Catalog Number: PT 000002 XXX XXX XX PT 000318 PT 000319 Distributed to: Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland

Z-1203-2021
Recall number
Z-1203-2021
Initiated
February 01, 2021
Classification
Class II
Status
Terminated
Recalling firm
Insulet Corporation
Quantity
21,723 guides

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Certain foreign user guides include a misprint which could lead to an incorrect bolus delivery and hypoglycemia. Only Outside the US (OUS): Error in the Omnipod DASH User Guide text, page 6 of the User Guide, the text states that the comma (,) is not used for the decimal on the PDM, when in fact it is the period (.) that is not used.If a user were to use the period as a decimal separator when entering a bolus value, this decimal separator would not be recognized by the device and any numbers entered after the period could result in a larger value than intended

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Certain foreign user guides include a misprint which could lead to an incorrect bolus delivery and hypoglycemia. Only Outside the US (OUS): Error in the Omnipod DASH User Guide text, page 6 of the User Guide, the text states that the comma (,) is not used for the decimal on the PDM, when in fact it is the period (.) that is not used.If a user were to use the period as a decimal separator when entering a bolus value, this decimal separator would not be recognized by the device and any numbers entered after the period could result in a larger value than intended

Code information

Part Number/Rev./Country: PT 000002 AUT DEU mg 1, 2, 3 Guide, Dash User, German Austria PT 000002 BEL DEU MG 1, 2, 3 Guide, Dash User Guide, Belgium Germany, mgdl PT 000002 BEL FRE MG 1, 2, 3, 4 Guide, Dash User Guide, Belgium French, mgdl PT 000002 BEL NLD MG 1, 2, 3 Guide, Dash User Guide, Belgium Dutch, mgdl PT 000002 DEU DEU mg 1, 2, 3 Guide, Dash User Guide, German Germany, mgdl PT 000002 DEU DEU mm 1, 2, 3 Guide, Dash User Guide, German Germany, mmoll PT 000002 DNK DANmm 1, 2, 3 Guide, Dash User Guide, Danish Denmark PT 000002 FIN FIN mm 1, 2, 3 Guide, Dash User Guide, Finnish Finland PT 000002 FRA FRE mg 1, 2, 3, 4 Guide, Dash User Guide, French France PT 000002 NLD NLD mm 1, 2, 3, 4 Guide, Dash User Guide, Dutch Netherland PT 000002 NOR NORmm 1, 2, 3 Guide, Dash User Guide, Norwegian Norway PT 000002 SWE SWEmm 1, 2, 3 Guide, Dash User Guide, Swedish Sweden 41333 CAN 1, 2 User Guide, Omnipod, Dash, Canada PT 000318 1, 2 Guide, Dash User Guide, Greece PT 000319 1, 2 Guide, Dash User Guide, Turkey

Distribution pattern

Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland