device · product 1 of 1
Dimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: used as an aid in the diagnosis of lipid disorders (such as diabetes mellitus), various liver and renal diseases, and in the assessment of risk for atherosclerosis and cardiovascular disease.
- Recall number
- Z-1188-2021
- Initiated
- February 03, 2021
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 4174 devices (Domestic: 2836 devices, Foreign: 1338 devices)
App-derived interpretation
Siemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to discolored reagent observed in some wells 5 and 6 of Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridges lot 20062BA.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Siemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to discolored reagent observed in some wells 5 and 6 of Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridges lot 20062BA.
Code information
Product: DV HDLC; Lot #: 20062BA; UDI Number: 0084276802043820062BA21030210464340840
Distribution pattern
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY, and the countries of Austria, Belgium, Canada, Denmark, France, Germany, Italy, Netherlands, Portugal, SA, Slovakia, Slovenia, Spain, Switzerland.