Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87427

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 22, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips North America Llc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Incisive CT Computed Tomography X-ray System Model: 728143

Z-1265-2021
Recall number
Z-1265-2021
Initiated
February 22, 2021
Classification
Class II
Status
Terminated
Recalling firm
Philips North America Llc
Quantity
105 units (1 US and 104 OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees

Code information

Serial Numbers: US: 500230 OUS: 500246 500260 500277 500261 530004 530102 530101 530103 530107 530109 530108 530113 530110 530119 530118 530117 530120 530125 530121 530124 530128 530126 530127 530129 530130 530132 530133 530134 530135 530136 530137 530139 530141 530138 530140 530144 530143 530145 530148 530146 530142 530147 530149 530154 530151 530158 530150 530153 530152 530157 530156 530159 530165 530161 530160 530162 530164 530163 530166 530171 530167 530168 530170 530169 530174 530173 530176 530185 530179 530172 530177 530196 530200 530201 530211 530212 530219 530221 500048 500231 500244 500247 500254 500257 500273 500270 500269 500226 500227 500241 500276 500229 500245 500248 500274 500263 500258 500278 500286 500225 500233 500256 500264 500265 UDI: (01)00884838059849(21)304034 (01)00884838059849(21)325078 (01)00884838085015(21)500218 (01)00884838085015(21)500219 (01)00884838085015(21)500221 (01)00884838085015(21)500222 (01)00884838085015(21)500223 (01)00884838085015(21)500224 (01)00884838085015(21)500225 (01)00884838085015(21)500226 (01)00884838085015(21)500227 (01)00884838085015(21)500229 (01)00884838085015(21)500231 (01)00884838085015(21)500233 (01)00884838085015(21)500234 (01)00884838085015(21)500237 (01)00884838085015(21)500238 (01)00884838085015(21)500239 (01)00884838085015(21)500241 (01)00884838085015(21)500242 (01)00884838085015(21)500245 (01)00884838085015(21)500247 (01)00884838085015(21)500248 (01)00884838085015(21)500249 (01)00884838085015(21)500251 (01)00884838085015(21)500252 (01)00884838085015(21)500254 (01)00884838085015(21)500255 (01)00884838085015(21)500256 (01)00884838085015(21)500257 (01)00884838085015(21)500258 (01)00884838085015(21)500261 (01)00884838085015(21)500264 (01)00884838085015(21)500274 (01)00884838085015(21)500283 (01)00884838085015(21)500296 (01)00884838085015(21)530038 (01)00884838085015(21)530107 (01)00884838085015(21)530108 (01)00884838085015(21)530109 (01)00884838085015(21)530113 (01)00884838085015(21)530117 (01)00884838085015(21)530118 (01)00884838085015(21)530119 (01)00884838085015(21)530120 (01)00884838085015(21)530121 (01)00884838085015(21)530124 (01)00884838085015(21)530125 (01)00884838085015(21)530126 (01)00884838085015(21)530127 (01)00884838085015(21)530128 (01)00884838085015(21)530129 (01)00884838085015(21)530130 (01)00884838085015(21)530132 (01)00884838085015(21)530133 (01)00884838085015(21)530134 (01)00884838085015(21)530135 (01)00884838085015(21)530136 (01)00884838085015(21)530137 (01)00884838085015(21)530138 (01)00884838085015(21)530139 (01)00884838085015(21)530140 (01)00884838085015(21)530141 (01)00884838085015(21)530142 (01)00884838085015(21)530143 (01)00884838085015(21)530144 (01)00884838085015(21)530145 (01)00884838085015(21)530146 (01)00884838085015(21)530147 (01)00884838085015(21)530148 (01)00884838085015(21)530149 (01)00884838085015(21)530150 (01)00884838085015(21)530151 (01)00884838085015(21)530152 (01)00884838085015(21)530153 (01)00884838085015(21)530154 (01)00884838085015(21)530156 (01)00884838085015(21)530157 (01)00884838085015(21)530158 (01)00884838085015(21)530159 (01)00884838085015(21)530160 (01)00884838085015(21)530161 (01)00884838085015(21)530162 (01)00884838085015(21)530163 (01)00884838085015(21)530164 (01)00884838085015(21)530165 (01)00884838085015(21)530166 (01)00884838085015(21)530167 (01)00884838085015(21)530168 (01)00884838085015(21)530169 (01)00884838085015(21)530170 (01)00884838085015(21)530171 (01)00884838085015(21)530172 (01)00884838085015(21)530173 (01)00884838085015(21)530174 (01)00884838085015(21)530176 (01)00884838085015(21)530177 (01)00884838085015(21)530179 (01)00884838085015(21)530180 (01)00884838085015(21)530185 (01)00884838085015(21)530196 (01)00884838085015(21)530200 (01)00884838085015(21)530211 (01)00884838085015(21)530219 (01)00884838085015(21)530221

Distribution pattern

US Virgin Island Foreign: Australia Bolivia China Germany Hungary Indonesia Italy Japan Korea Korea, Republic of Latvia Palestine Philippines Romania Russian Federation Senegal Thailand Uzbekistan