device · product 1 of 1
CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Recall number
- Z-1272-2021
- Initiated
- February 26, 2021
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Accuray Incorporated
- Quantity
- 243 units
App-derived interpretation
The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.
Official device-enrichment evidence · Sourced
Equipment maintenance
Inspect official wording and provenance
Reason for recall
The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.
Code information
Product Codes: 019200, 020600, 021831, 022986, 023730, 025000, 027600, 028000, 030000, 031000, 031500, 0660000, 029640-001, 029828-003, 031000-004, 032000-001, 032000-004, 033000-000, 033000-010, 053101-010, 053101-M-000, 053301-010, 053301-M-000, 053401-010, 053401-M-000, 054000-002, 054000-004, 0660000-000, 1010140-000, 1047641-000 Serial Numbers: C0063, C0067, C0008, C0076, C0077, C0085, C0098, C0101, C0112, C0126, C0131, C0135, C0138, C0151, C0155, C0162, C0166, C0172, C0174, C0176, C0177, C0186, C0187, C0188, C0195, C0197, C0200, C0206, C0208, C0211, C0213, C0228, C0231, C0234, C0240, C0242, C0246, C0253, C0263, C0266, C0271, C0239, C0441, C0446, C0447, C0456, C0460, C0464, C0465, C0469, C0472, C0475, C0481, C0506, C0507, C0193, C0241, C0293, C0090, C0092, C0093, C0100, C0103, C0104, C0115, C0117, C0119, C0124, C0130, C0148, C0150, C0153, C0168, C0181, C0182, C0194, C0205, C0226, C0235, C0236, C0244, C0254, C0255, C0258, C0259, C0267, C0268, C0274, C0276, C0280, C0282, C0283, C0285, C0286, C0289, C0291, C0292, C0297, C0298, C0299, C0300, C0301, C0302, C0303, C0304, C0305, C0306, C0309, C0310, C0312, C0313, C0317, C0319, C0321, C0325, C0328, C0329, C0331, C0339, C0347, C0353, C0357, C0365, C0377, C0382, C0386, C0388, C0399, C0401, C0403, C0405, C0416, C0432, C0435, C0451, C0064, C0071, C0074, C0091, C0099, C0102, C0137, C0238, C0249, C0252, C0270, C0277, C0294, C0417, C0476, C0384, C0414, C0424, C0439, C0427, C0428, C0344, C0423, C0372, C0387, C0330, C0343, C0345, C0346, C0348, C0351, C0352, C0356, C0360, C0363, C0368, C0369, C0370, C0373, C0380, C0381, C0383, C0389, C0390, C0394, C0395, C0396, C0397, C0398, C0402, C0404, C0406, C0411, C0413, C0415, C0420, C0422, C0429, C0430, C0436, C0437, C0438, C0440, C0444, C0445, C0448, C0453, C0457, C0458, C0466, C0461, C0467, C0473, C0480, C0482, C0484, C0486, C0487, C0488, C0489, C0490, C0491, C0492, C0493, C0494, C0495, C0496, C0497, C0500, C0505, C0513, C0014, C0022, C0040, C0042, C0046, C0047, C0056, C0059, C0066, C0079, C0086, C0089, C0140, C0173, C0311, C0355, C0470
Distribution pattern
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ. CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV, PR and the countries of Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Republic of Korea, Latvia, Luxembourg, Malaysia, Mexico, Oman, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.