Exact FDA event identifier

Event 87662

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 06, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

Implantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Azure family: Azure XT DR MRI SureScan", Azure S DR MRI SureScan, Azure XT SR MRI SureScan, Azure S SR MRI SureScan. 2. Astra family: Astra XT DR MRI SureScan, Astra XT SR MRI SureScan.

Z-1525-2021
Recall number
Z-1525-2021
Initiated
April 06, 2021
Classification
Class II
Status
Ongoing
Quantity
575815 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Code information

all codes

Distribution pattern

Worldwide distribution.

device · product 2 of 4

Implantable cardiac resynchronization therapy pacemakers (CRT-P) - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Percepta family: Percepta CRT-P MRI SureScan, Percepta Quad CRT-P MRI SureScan. 2. Serena family: Serena CRT-P MRI SureScan, Serena Quad CRT-P MRI SureScan. 3. Solara family: Solara CRT-P MRI SureScan, Solara Quad CRT-P MRI SureScan.

Z-1526-2021
Recall number
Z-1526-2021
Initiated
April 06, 2021
Classification
Class II
Status
Ongoing
Quantity
333408 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Code information

all codes

Distribution pattern

Worldwide distribution.

device · product 3 of 4

Programmers - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. 2090 Carelink 2. 29901 Encore

Z-1527-2021
Recall number
Z-1527-2021
Initiated
April 06, 2021
Classification
Class II
Status
Ongoing
Quantity
37690 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Code information

all codes

Distribution pattern

Worldwide distribution.

device · product 4 of 4

CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A

Z-1528-2021
Recall number
Z-1528-2021
Initiated
April 06, 2021
Classification
Class II
Status
Ongoing
Quantity
16,311 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Code information

all codes

Distribution pattern

Worldwide distribution.