Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87854

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 26, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Roche Diagnostics Operations, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

IRON GEN 2: ROCHE COBAS INTEGRA ANALYZER, REAGENT CASSETTES, Catalog#: 03183696122 - Product Usage: In vitro test for the quantitative determination of iron in human serum and plasma on Roche/Hitachi cobas c and cobas Integra systems.

Z-1675-2021
Recall number
Z-1675-2021
Initiated
April 26, 2021
Classification
Class II
Status
Ongoing
Quantity
20,468 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Roche received several complaints regarding increased control recovery and discrepant elevated results for the Iron Gen.2 (IRON2) assay on the cobas c311, 501, and 502 analyzers and on the COBAS Integra 400 plus analyzer.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Roche received several complaints regarding increased control recovery and discrepant elevated results for the Iron Gen.2 (IRON2) assay on the cobas c311, 501, and 502 analyzers and on the COBAS Integra 400 plus analyzer.

Code information

UDI: 04015630918904

Distribution pattern

US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.